Overview
The function and range settings are always correct because the measurement parameters are detected and set automatically. Three easy-to-read buttons replace the traditional rotary switch when selecting manual test functions.
Additional benefits:
APO (Auto Power Off) saves battery life
Overmold housing makes it rugged
Large, backlit, dual readout display with analog bar graph is easy-to-read
Lead holder conveniently molded into housing for easy storing
Testo’s history of efficiency, innovation, and quality now extends into the entire electrical lineup. Once again, Testo brings high value, feature-packed products to the HVAC contractor market.
What's included:
testo 760-3 digital multimeter 1000 V with TRMS, capacitance, duty cycle, low pass filter, auto-test setup, and temperature measuring
includes 3 AAA batteries, test leads
Technical data | |
---|---|
Measuring range |
-4.0° to 932.0 °F / -20 to +500 °C |
Accuracy |
±1.8 °F (32° to 212 °F) / ±1 °C (0 to +100 °C) |
Resolution |
0.1 °F / 0.1 °C |
Technical data | |
---|---|
Resolution |
600 mV / 0.1 mV |
Accuracy |
± (0.8 % of mv + 3 Digit) |
Technical data | |
---|---|
Resolution |
600 mV / 0.1 mV |
Accuracy |
± (0.8 % of mv + 3 Digit) |
Technical data | |
---|---|
Resolution |
600 μA / 0.1 μA |
Accuracy |
± (1.5 % of mv + 5 Digit) |
Technical data | |
---|---|
Resolution |
600 μA / 0.1 μA |
Accuracy |
± (1.5 % of mv + 5 Digit) |
Technical data | |
---|---|
Resolution |
60.00 Ohm / 0.01 Ohm |
Accuracy |
± (1.5 % of mv + 3 Digit) |
Technical data | |
---|---|
Resolution |
6.000 nF / 0.001 nF |
Accuracy |
± (0.5 % of mv + 3 Digit) |
Technical data | |
---|---|
Resolution |
600.0 Hz / 0.1 Hz |
Accuracy |
± (0.1 % of mv + 1 Digit) |
Technical data | |
---|---|
Protection class |
IP64 |
Warranty |
2 years |
Storage temperature |
5° to 140 °F / -15 to +60 °C |
Weight |
12 oz. / 340 g |
Overvoltage Category |
CAT IV 600V; CAT III 1000V |
Authorizations |
TÜV; CSA; CE |
Dimensions |
7 x 3 x 2 in. / 167 x 85 x 45 mm |
Operating temperature |
14° to 122 °F / -10 to +50 °C |
Technical data | |
---|---|
Diode test |
yes |
Continuity Testing |
yes |
Current probe measurement with adapter |
yes |
Display ligthing |
yes |
Hold |
yes |
MinMax |
yes |
Temperature measurement with adapter |
yes |
True RMS |
yes |
Low-pass filter |
yes |
Duty cycle measurement |
yes |
Display (counts) |
6,000 |