Testo 340 - Flue gas analyzer for industry emission measurement

 
CODE: 0632 3340 S2
 
Emission checks in industry are easy with the flue gas analyzer testo 340. Professional technology meets practical design
 
Disponibilitate: In stoc
 
 
 

Product description

Precise technology: That’s what you can rely on when you’re working out in the field. But precise technology is nothing if it is not versatile, easy to use and adaptable for use in the field. The testo 340 flue gas analyzer meets all these expectations and much more.
Industrial emission measurement with testo 340 flue gas analyzer
Which sensors do you need in the course of your daily work and which measurements do you normally need to carry out? The testo 340 flue gas analyzer comes standard-fitted with an O2 sensor and four more sensors (CO, COlow, NO, NOlow, NO2 or SO2 ) for you to choose from, allowing you to customize your flue gas analyzer to suit your individual demands. The analyzer’s compact design and the reliable technology ensure increased mobility, making it ideal for use in on-site service operations and for installing or carrying out checks on firing plants or energy production systems. To find out more about the different application possibilities and to see which industrial applications your testo 340 flue gas analyzer is suitable for, just click on the “Applications” tab.


High gas concentrations? No problem with the flue gas analyzer testo 340!
The flue gas analyzer has an extended measuring range which allows you to carry out measurements even when gas concentrations are very high (i.e. during industrial flue gas analysis). The extended measuring range (thinning) is activated automatically when high gas concentrations suddenly occur. The demands on the sensor are no different to than when the flue gas analyzer is used in low gas concentrations.

What's included

testo 340 flue gas analyzer, calibration certificate, shoulder strap, O2 sensor and integrated flow and pressure difference measurements.
Please note that your testo 340 must be fitted with a second gas sensor for it to work. Up to three additional sensors can be fitted.


 

Differential Pressure - Piezoresistive

Technical data  

Measuring range

-200 to 200 hPa

Accuracy

±0.5 hPa (-49.9 to 49.9 hPa)
±1.5 % of mv (Remaining Range)

Resolution

0.1 hPa 

 

Absolute Pressure

Technical data  

Measuring range

+600 to +1150 hPa

Accuracy

±10 hPa

Resolution

1 hPa

 

Flue gas O₂

Technical data  

Measuring range

0 to 25 Vol.%

Accuracy

±0.2 Vol.%

Resolution

0.01 Vol.%

Reaction time t₉₀

< 20 s

 

Flue gas CO (with H₂-compensation)

Technical data  

Measuring range

0 to 10000 ppm

Accuracy

±10 ppm or ±10 % of mv (0 to 200 ppm)
±20 ppm or ±5 % of mv (201 to 2000 ppm)
±10 % of mv (2001 to 10000 ppm)

Resolution

1 ppm

Reaction time t₉₀

< 40 s

 

Flue gas COlow

Technical data  

Measuring range

0 to 500 ppm

Accuracy

±2 ppm (0 to 39.9 ppm)
±5 % of mv (Remaining Range) data corresponds to 20°C ambient temperature. Additional temperature coefficient 0.25% of reading/K.

Resolution

0.1 ppm

Reaction time t₉₀

< 40 s 

 

Flue gas NO

Technical data  

Measuring range

0 to 4000 ppm

Accuracy

±5 ppm (0 to 99 ppm)
±5 % of mv (100 to 1999 ppm)
±10 % of mv (2000 to 4000 ppm)

Resolution

1 ppm

Reaction time t₉₀

< 30 s

 

Flue gas Nolow

Technical data  

Measuring range

0 to 300 ppm

Accuracy

±2 ppm (0 to 39.9 ppm)
±5 % of mv (Remaining Range)

Resolution

0.1 ppm

Reaction time t₉₀

< 30 s

 

Flue gas NO₂

Technical data  

Measuring range

0 to 500 ppm

Accuracy

±10 ppm (0 to 199 ppm)
±5 % of mv (Remaining Range)

Resolution

0.1 ppm

Reaction time t₉₀

< 40 s

To avoid absorption, a maximum measurement duration of 2 hours should not be exceeded.

 

Flue gas SO₂

Technical data  

Measuring range

0 to 5000 ppm

Accuracy

±10 ppm (0 to 99 ppm)
±10 % of mv (Remaining Range)

Resolution

1 ppm

Reaction time t₉₀

< 40 s

To avoid absorption, a maximum measurement duration of 2 hours should not be exceeded.

 

Flue gas Draught

Technical data  

Measuring range

-40 to +40 hPa

Accuracy

±0.03 hPa (-2.99 to +2.99 hPa)
±1.5 % of mv (Remaining Range)

Resolution

0.01 hPa

 

Temperature

Technical data  

Measuring range

-40 to +1200 °C

Accuracy

±0.5 °C (0 to +99 °C)
±0.5 % of mv (Remaining Range)

Resolution

0.1 °C

 

Flue gas degree of effectivity, Eta (calculated)

Technical data  

Measuring range

0 to 120 %

Resolution

0.1 %

 

Flue gas loss (calculated)

Technical data  

Measuring range

0 to 99.9 %

Resolution

0.1 %

 

Flue gas dewpoint (calculated)

Technical data  

Measuring range

0 to 99.9 °Ctd

Resolution

0.1 °Ctd

 

Flue gas CO₂ calculation (calculated from O₂)

Technical data  

Measuring range

0 to CO₂ max

Accuracy

±0.2 Vol.%

Resolution

0.1 Vol.%

Reaction time t₉₀

< 40 s

 

General technical data

Technical data  

Housing

TPE PC

Protection class

IP40

Warranty

Measuring instrument 2 years (excepting wearing parts e.g. gas sensors, etc.); rechargeable battery: 12 months; CO sensor: 12 months; Sensor NOlow: 12 months; Sensor NO2: 12 months; Sensor SO₂: 12 months; Sensor O₂: 18 months; NO sensor 12 months; COlow sensor 12 months

Display size

160 x 240 pixels

Display funktion

Graphic display

Power supply

Battery block 3.7 V / 2.4 Ah, Mains unit 6.3 V / 2 A

 